What it does
Heat kills microbes faster the hotter it gets. The F0 value puts a whole autoclave cycle, heating up, holding and cooling down, on one scale: the number of minutes at 121.1 °C that would kill as much. This page works it out from your own temperature log:
- Lethal rate. Each reading counts as L = 10^((T − Tref) / z) minutes per minute. At the reference temperature (121.1 °C) L is 1; 10 °C hotter it is 10; 10 °C cooler, 0.1. The z-value (10 °C for steam by convention) is how many degrees make the kill ten times faster.
- F0. The lethal rates are added up over time, with straight lines between readings (trapezoids) or each reading held until the next, as some loggers do. You see the total and a running F0 curve.
- Probes. Every temperature column is a probe. The page shows F0 for each and picks the coldest one (lowest F0), because the slowest part of the load decides whether the load is sterile.
- Hold. Time at or above the hold temperature, when the hold started and ended, and the F0 gathered during it.
- Equivalent time at another temperature, such as 134 °C.
- Kill of one organism. Type a D-value (minutes for a 90 % kill at a stated temperature) and the organism’s own z. The page works out its log reduction over your cycle, the organisms left per item, whether you reach a target sterility assurance level (SAL) and how long a steady hold at any temperature would need.
How to use it
- Log. Paste two or more columns: time, then a temperature for each probe. Minutes, seconds, hours and clock times (10:15:30) all work, and °F is converted. You can also open a CSV or Excel file.
- Settings. Keep 121.1 °C and z = 10 °C for steam, or type your own (dry heat is often worked out with a different reference and z). Set the hold temperature.
- Read F0 for each probe, look at the curves and download the CSV.
- Organism. Type the D-value, its temperature and z from your biological indicator’s certificate, the starting number per item and the target SAL. Download a PNG, PDF or Word report.
Worked example
The example is a made-up 45-minute cycle with a reading every minute: a chamber probe and a probe in the centre of the load that heats up more slowly. With Tref = 121.1 °C and z = 10 °C:
- The chamber reaches F0 = 27.65 min; the load centre only 22.10 min, so it is the coldest probe.
- The load centre is at or above 121 °C for 15.45 min, from 16.60 to 32.05 min, and gathers 16.58 min of F0 there; the rest comes from heating up and cooling down.
- 22.10 min at 121.1 °C kills as much as 1.134 min at 134 °C.
For the made-up organism (D = 1.5 min at 121.1 °C, z = 10 °C) the load centre gives a 14.73 log reduction. Starting from 10⁶ organisms per item, a SAL of 10⁻⁶ needs 12 logs, so the cycle meets the target with 2.735 logs to spare and leaves 1.84 × 10⁻⁹ organisms per item. A steady hold for 12 logs would be 18 min at 121.1 °C, or 0.9232 min at 134 °C.
Tips
- Put the probe where the load heats slowest (the middle of a dense pack or the largest bottle) and judge the cycle by that probe.
- Log at least once a minute; with gaps, F0 is only a straight-line guess and the page warns you.
- F0 assumes saturated steam. Trapped air means the same temperature kills less, so F0 does not replace a Bowie-Dick test or biological indicators.
FAQ
Why is F0 above the hold time? The chamber sat a little above 121.1 °C, and heating up and cooling down add lethality too.
Which z should I use for the organism? Its own, from the indicator’s certificate. The F0 z (10 °C) is a convention; the organism z changes its log reduction.
Is this a validation? No. It is a calculator to check and explain numbers. Follow your own validated procedure and the official standards.
Where the numbers come from
- The lethal rate L = 10^((T − Tref) / z), F0 as the time integral of L, D-value shifts D_T = D_ref × 10^((T_ref − T) / z), log reduction = F ÷ D and the sterility assurance level are general sterilisation science, written by fibub.
- Checked in the tests against an independent NumPy script on the example log and three random logs with random reference temperatures and z-values (F0 totals, running totals, time at the hold temperature, a second z), against the exact integral of a straight-line ramp, and against hand working for the D-value and hold-time arithmetic.
- The example autoclave log and the example organism (D = 1.5 min at 121.1 °C, z = 10 °C) are made up for this tool. All values are editable.
- A log you paste or open is read on the page only. It is never uploaded or stored.