A copper sensor's 28.05 mV slope: activity or the electrode?
A 2025 solid-contact copper electrode reads 28.05 mV per tenfold change instead of the ideal 29.58. We rebuild the Nernst line and show that most of the shortfall is what activity coefficients do to a concentration calibration.
An ion-selective electrode is judged almost entirely by one number: how many millivolts its potential moves for every tenfold change in the ion it is built for. The Nernst equation fixes that number exactly, and for a doubly charged ion such as Cu²⁺ at 25 °C it is 29.58 mV. A 2025 paper in BMC Chemistry builds a solid-contact copper electrode (a membrane electrode with no inner filling solution, using polyaniline nanoparticles to carry the signal to the wire) and measures 28.05 mV per decade — which the authors fairly call near-Nernstian. This article uses the Nernst calculator to ask a sharper question: how much of that 1.5 mV shortfall is the electrode, and how much is simply what happens when you calibrate against concentration instead of activity? Along the way the calibration curve handles the straight-line fit and its detection limit, and cyclic voltammetry analyser is the tool for the other half of an electrode-characterisation paper.
The paper in brief
The electrode is a PVC membrane holding β-cyclodextrin and Reinecke’s salt, coated onto a solid contact of polyaniline nanoparticles. The authors report a response over six decades, from 10⁻⁸ to 10⁻² mol/L Cu²⁺, a slope of 28.05 mV per decade, a detection limit of 7.5 × 10⁻⁹ mol/L, a response time of 10 s or less and a stable reading between pH 6 and 9. They then use the sensor for something unusual: watching copper nanoparticles form and dissolve in real time, which needs a fast electrode rather than a sensitive one, and measuring copper in spiked water with 95.8 to 98.9 % recovery.
The detection limit is worth converting before anything else. At copper’s molar mass of 63.546 g/mol, 7.5 × 10⁻⁹ mol/L is 0.48 µg/L: the electrode sees copper roughly a thousand times below the milligram-per-litre levels where drinking-water rules sit (check your own country’s official limit, which changes).
Checked with the tools
Step 1: what the ideal slope is, to four figures. The Nernst slope is RT ln10 / zF. The tool’s constants give 29.5797 mV per decade for z = 2 at 25.00 °C (and 59.159 mV for a singly charged ion, the familiar number). The paper’s 28.05 is 94.8 % of it.
Step 2: can temperature explain the gap? No. The slope rises by only 0.099 mV per decade per °C: 29.084 mV at 20 °C, 30.076 mV at 30 °C. To reach 28.05 mV per decade by temperature alone, the solution would have to sit at about 9.6 °C (Figure 2). So a lab at 20 to 30 °C can move the slope by half a millivolt, not one and a half.
Step 3: activity, the usual culprit. The Nernst equation is written in activity, the effective concentration once ions get in each other’s way, and activity falls below concentration as a solution gets saltier. For Cu²⁺ in a CuCl₂ solution the ionic strength is three times the concentration, and the extended Debye-Hückel expression gives an activity coefficient of 0.52 at 10⁻² mol/L, 0.92 at 10⁻⁴ and 0.99 at 10⁻⁶. So the top decade of the calibration is pulled down while the bottom decades are untouched, and a straight line fitted to potential against concentration across those six decades has a slope of only 28.62 mV per decade even for a perfect electrode.
That accounts for about two-thirds of the gap between 29.58 and the measured 28.05. Figure 1 draws all three lines together, anchored to agree in the middle of the range: they differ by less than 5 mV anywhere in six decades, which is the real message. The electrode is behaving close to thermodynamically; the slope label alone cannot tell you otherwise unless the calibration states whether it is in activity or in concentration, and at what ionic strength.
Step 4: what a 1.5 mV slope error costs in a result. If you calibrate at one concentration and then read a sample some decades away, using 29.58 mV where the electrode really gives 28.05 mV multiplies your answer by 10^(1.53/29.58) per decade, that is 1.13× per decade and 2.04× over six decades. Over the whole published range the electrode’s own span is 168.3 mV, against 177.5 mV for the ideal line. So: calibrate with standards that bracket your samples, and the slope question almost disappears; extrapolate six decades and it doubles your answer.
| Check | Paper | Tool result |
|---|---|---|
| Ideal slope, Cu²⁺ at 25 °C | 28.05 mV/decade measured, “near-Nernstian” | 29.5797 mV/decade (94.8 % reached) |
| Temperature needed for 28.05 mV/decade | measured at room temperature | about 9.6 °C |
| Slope against concentration with real activities | not stated | 28.62 mV/decade for a perfect electrode |
| Potential span over 10⁻⁸ to 10⁻² mol/L | 6 decades of response | 177.5 mV ideal, 168.3 mV at the measured slope |
| Detection limit as a mass | 7.5 × 10⁻⁹ mol/L | 0.48 µg/L of copper |
| Cu²⁺/Cu potential at 10⁻² and 10⁻⁸ mol/L | not reported | +0.283 V and +0.105 V vs SHE (+0.086 and −0.092 V vs Ag/AgCl, saturated KCl) |
How we made the graphs
The charts and every number come from scripts/blog-figures/solid-contact-copper-ion-selective-electrode-nernstian-slope.mjs in fibub’s public code. It calls thermalV, halfE, half, convertRef and qFromE from src/lib/sci/nernst.ts, the maths behind the Nernst calculator, plus logGamma from src/lib/sci/buffer.ts (the activity coefficient used by the buffer recipe calculator) and molarMass from src/lib/sci/chem.ts.
The slope, the range, the detection limit and the pH window come from the paper’s abstract. The standard potential of Cu²⁺/Cu (+0.3419 V vs SHE) comes from the tool’s own table. Our assumptions are the chemistry behind the activity calculation: that the calibration standards were CuCl₂ in water with no added background salt, so the ionic strength is three times the copper concentration, and that the three lines should be anchored to agree at 10⁻⁵ mol/L. The paper’s own calibration points are not reproduced here; nothing in these figures is its data.
How we test the maths
The Nernst maths is pinned in tests/sci.mjs, which runs on every change to the site:
- The Daniell cell: E°cell must be +1.1037 V, and with 0.01 mol/L Cu²⁺ the cell must read +1.0445 V.
- The slope itself: RT ln10 / F at 25 °C must be 59.159 mV, and every one of the 44 standard half-reactions must balance for charge and atoms.
- Reference conversions: 0.500 V against SCE must become 0.544 V against Ag/AgCl (saturated KCl), and a potential against Ag/AgCl converted to RHE at pH 13 must shift by 0.197 + 0.769 V.
- New with this article, in
tests/sci-tools/nernst-calculator.mjs: the divalent slope at 25 °C must be 29.5797 mV/decade and change by 0.0992 mV per °C; a Cu²⁺/Cu electrode must read +0.2827 V at 10⁻² mol/L and +0.1053 V at 10⁻⁸ mol/L vs SHE; and the straight line through potential against log concentration with Debye-Hückel activities over those six decades must come out at 28.62 mV/decade.
Check it yourself
On the Nernst calculator:
- In the Electrode or cell card, press One electrode.
- Choose +0.342 V Cu²⁺(aq) + 2e⁻ → Cu(s) as the electrode and leave Temperature (°C) at 25.
- Under Concentrations, set the Cu²⁺ box to 0.01 (the tool treats the number you type as an activity). The result line reads E = +0.2827 V vs SHE, and the facts underneath give the same potential against Ag/AgCl and SCE, with the Nernst slope for this half-reaction.
- Change the activity to 1e-8: the potential falls to +0.1053 V. The two readings differ by 177.5 mV, which is six decades at 29.58 mV each.
- Set Temperature (°C) to 9.6 and look at the slope the tool reports: 28.05 mV per decade, the paper’s number, which is how you can see that temperature is not where it came from.
- In the Reference electrode converter card, type 0.2827 in Potential (V), set Measured against to SHE / NHE and read the table: +0.086 V against Ag/AgCl (saturated KCl) and +0.042 V against SCE.
- In Find Q or a concentration from a measured potential, put 0.194 in Measured E (V), 0.3419 in E° (V, same reference) and 2 in Electrons n: Q comes back as about 1.0 × 10⁵, which is 1/activity for 10⁻⁵ mol/L Cu²⁺.
What researchers can use it for
- Grade a new sensor honestly: work out the ideal slope at your temperature and your ion’s charge before calling a slope sub-Nernstian.
- Decide whether to fix the ionic strength: compute the activity coefficient at the top of your range; if it costs you more than a millivolt per decade, add a constant background salt and say so in the paper.
- Set the calibration range: pick standards that bracket the samples, and use the slope-error numbers above to show what extrapolation would cost.
- Convert potentials between references when comparing papers, including to RHE at a given pH, so Ag/AgCl and SCE numbers can be read side by side.
- Turn a measured potential back into a concentration (the Q box) when a calibration has drifted, and fit the calibration itself, with its detection limit, in the calibration curve.
The fine print
- We worked from the paper’s abstract: its figures, calibration table and selectivity coefficients were not available to us, so nothing here is a fit to its measurements.
- The activity calculation assumes CuCl₂ in water with no background electrolyte. If the authors fixed the ionic strength, activity explains less of the slope and the electrode’s own response explains more.
- The extended Debye-Hückel expression is reliable to roughly 0.1 mol/L ionic strength; 10⁻² mol/L CuCl₂ is inside that, but only just.
- A real membrane electrode’s slope also drifts with ageing, with the film thickness and with interfering ions. A thermodynamic check cannot separate those; it only sets the ceiling.
- The Cu²⁺/Cu potentials here are for a copper metal electrode, shown as the thermodynamic yardstick. The paper’s device is a membrane electrode, whose standard potential depends on its own chemistry.
Sources
- Eid, S. M., Elsherbiny, M. S., Abbas, M. M., Abdel Ghany, M. F. and Kelani, K. M. (2025), A nanostructured solid-contact electrode for real-time monitoring of copper nanoparticle dynamics and environmental analysis, BMC Chemistry 20, doi:10.1186/s13065-025-01692-y. Licence CC BY 4.0 (Creative Commons Attribution), checked with Crossref metadata for the DOI, 2026-11-24. Values used here, all from the paper's abstract: polyaniline-nanoparticle solid contact under a PVC membrane with beta-cyclodextrin and Reinecke's salt; slope 28.05 mV per decade (described as near-Nernstian); linear range 1 x 10-8 to 1 x 10-2 mol/L; detection limit 7.5 x 10-9 mol/L; response time 10 s or less; working pH 6 to 9; activation energy 34.06 kJ/mol; recovery 95.8 to 98.9 % in spiked water.
- Standard reduction potential of Cu2+/Cu (+0.3419 V vs SHE) and reference-electrode potentials: the CRC Handbook electrochemical series, as used by fibub's Nernst calculator. Physical constants: CODATA (NIST).
- Activity coefficients: the extended Debye-Huckel expression used by fibub's Buffer recipe calculator, log gamma = -A z^2 (sqrt(I)/(1 + sqrt(I)) - 0.3 I).