Famotidine by charge-transfer colour: checking the absorbance window

A 2024 study measures famotidine as a coloured complex at 546 nm. We rebuild its Beer-Lambert line, confirm its molar absorptivity and Sandell sensitivity agree, and show why a longer cell would help.

Famotidine is a common stomach-acid tablet, and it has no strong colour of its own. A 2024 paper in Scientific Reports gets around that in the classic way: mix the drug with an electron-poor reagent, ortho-chloranil, let the two form a charge-transfer complex (a loosely bound pair where electron density shifts from one molecule to the other, which makes it absorb visible light), and read the colour at 546 nm. The paper reports a molar absorptivity, a Sandell sensitivity, a linear range and detection limits. Every one of those can be checked in a browser with the Beer–Lambert calculator, and that is what this article does. For the steps on either side, the calibration curve fits the standards and gives the detection limit from them, and method validation statistics turns the recovery and repeatability numbers into the usual validation statistics.

The paper in brief

The authors add ortho-chloranil to famotidine in an alkaline aqueous medium, wait for the colour to form and measure it against a water blank at 546 nm. They report that Beer’s law holds from 2 to 28 µg/mL with R² = 0.9977, a molar absorptivity (ε) of 2,159.648 L/(mol·cm), a Sandell sensitivity of 0.15625 µg/cm², a limit of detection of 0.3024 µg/mL and a limit of quantification of 1.471 µg/mL. Job’s method and the molar-ratio method both point to a 2:1 reagent-to-drug complex, and recovery from tablets is about 99.9 %. The method needs no organic solvent in the measurement step, which is its selling point: it is a cheap, green assay for a cheap drug.

Two of those numbers are independent of the instrument, so anyone can test them: ε and the Sandell sensitivity. The rest say where the method is comfortable, and that is where a Beer-Lambert check earns its keep.

Checked with the tools

Step 1: do ε and the Sandell sensitivity agree? The Sandell sensitivity is the mass of analyte per square centimetre of light path that gives an absorbance of 0.001, and for a Beer-Lambert absorber it is simply the molar mass divided by ε. Famotidine is C₈H₁₅N₇O₂S₃, which the molar mass calculator puts at 337.435 g/mol. The paper’s two numbers multiply to 0.15625 × 2,159.648 = 337.445 g/mol. The pair is self-consistent to 0.003 %, so the two sensitivity figures were worked out from the same calibration and the same molar mass. That is a check worth running on any spectrophotometric paper: when ε and the Sandell value do not multiply to the analyte’s molar mass, one of them is wrong.

Step 2: where does the published range sit on the absorbance scale? With ε = 2,159.648 L/(mol·cm) and the usual 1 cm cuvette, the calibration line has a slope of 0.0064 absorbance units per µg/mL. So the paper’s range of 2 to 28 µg/mL is a range of absorbance from 0.0128 to 0.1792. The whole calibration lives in the bottom tenth of the scale. The tool’s own reading-quality note calls everything under A = 0.05 “very low: noise and the blank matter a lot”, which covers every standard below 7.8 µg/mL, more than a fifth of the stated range.

Figure 1 shows the line, the strip the paper calibrated over, and the 0.1 to 1 absorbance window that an ordinary spectrophotometer measures most accurately. A 5 cm cell, the cheapest change possible, lifts the same concentrations to A = 0.064 to 0.896, almost exactly into that window.

Famotidine complex: absorbance against concentrationFamotidine complex: absorbance against concentrationthe paper's calibration range00.250.50.751051015202530Famotidine (µg/mL)Absorbance at 546 nmA = 1: top of the accurate rangeA = 0.1: bottom of it1 cm cell (the usual cuvette)5 cm cellfibub.com
Figure 1. The Beer-Lambert line for the published method (ε = 2,159.6 L/(mol·cm) at 546 nm), worked out with the Beer-Lambert calculator’s maths. The shaded strip is the paper’s 2 to 28 µg/mL calibration range; the dashed lines are the absorbance window an ordinary spectrophotometer measures best. Our figure, not the paper’s.

Step 3: what the low absorbance costs. Suppose the instrument reads absorbance to within ±0.002 A, which is ordinary for a routine single-beam spectrophotometer (our assumption; the paper does not state its photometric noise). That error carries straight into the result as 0.002 ÷ A. At the bottom of the published range it is 15.6 %; at the top, 1.1 %. In a 5 cm cell the worst case drops to 3.1 % (Figure 2). The paper’s own repeatability is far better than 15 %, which tells you its instrument is better than our assumed 0.002 A, or that the low standards were not the ones repeated.

What a ±0.002 absorbance error costsWhat a ±0.002 absorbance error costs05101551015202530Famotidine (µg/mL)Error in the result (%)2 %1 cm cell5 cm cellfibub.com
Figure 2. The same method seen as error: a fixed ±0.002 A reading error turned into a percentage of the result. Our assumption is the 0.002 A; the curve shape is just 1/A.

Step 4: the detection limit, read as an absorbance. A limit of detection of 0.3024 µg/mL means a signal of 0.0019 A in a 1 cm cell, about two thousandths of full scale. Working backwards through the usual 3.3σ/slope rule, that limit needs the blank’s standard deviation to be 0.00059 A. If the blank instead wobbles by 0.002 A, the same rule gives a detection limit of 1.03 µg/mL, three times higher. Note too that the paper’s quantification limit is 4.86 times its detection limit, while the conventional 10σ and 3.3σ definitions give a ratio of 3.03 — so the two limits come from different rules, and the paper says the detection limit came from the absorbance of its lowest standard rather than from blank statistics.

Check Paper Tool result
Sandell × ε (should be the molar mass) 0.15625 µg/cm² × 2,159.648 337.445 g/mol, against 337.435 for C₈H₁₅N₇O₂S₃
Absorbance of the lowest standard, 1 cm not stated 0.0128 (flagged “very low”)
Absorbance of the highest standard, 1 cm not stated 0.1792
Concentrations that give A = 0.1 to 1 at 1 cm 2 to 28 µg/mL calibrated 15.6 to 156 µg/mL
Same range in a 5 cm cell not tried A = 0.064 to 0.896
Blank noise implied by the stated LOD LOD 0.3024 µg/mL σ = 0.00059 A
LOQ ÷ LOD 4.86 3.03 for the 10σ/3.3σ convention

None of this says the assay is wrong. It says the assay is running on a tenth of the dynamic range it could use, and that a longer cell (or a more concentrated sample) would buy accuracy for nothing.

How we made the graphs

The charts and every number in them come from scripts/blog-figures/famotidine-charge-transfer-complex-molar-absorptivity.mjs in fibub’s public code. It calls solveBL and absQuality from src/lib/sci/beer.ts, the maths behind the Beer–Lambert calculator, and parseFormula with molarMass from src/lib/sci/chem.ts, which powers the molar mass calculator.

solveBL fills in whichever one of A, ε, path length and concentration is left empty, so the script uses it in both directions: forwards to turn each concentration into an absorbance, and backwards to find the concentration that gives A = 0.1 or A = 1. The inputs taken from the paper are ε, the Sandell sensitivity, the 2 to 28 µg/mL range, the detection and quantification limits and the 546 nm wavelength. The 1 cm and 5 cm path lengths, and the ±0.002 A reading error, are our assumptions. Nothing here is a fit to the paper’s measurements; the paper’s calibration points are only in its figures.

How we test the maths

Beer’s law is one multiplication, so the test is that the tool never slips a unit:

  • tests/sci-tools/beer-lambert-calculator.mjs pins the textbook NADH reading: ε = 6,220 L/(mol·cm) at 340 nm, 0.1 mmol/L in a 1 cm cell must give A = 0.622, and 10 % transmittance must come back as A = 1 exactly.
  • It checks the four ways round: given any three of A, ε, l and c, the fourth must come back, and asking for two at once must raise an error rather than guess.
  • New with this article: famotidine’s molar mass from C₈H₁₅N₇O₂S₃ must be 337.435 g/mol, the Sandell sensitivity 0.15625 µg/cm² times ε 2,159.648 must come back to that molar mass within 0.05 %, 2 µg/mL must give A = 0.01280 (flagged warn) and 28 µg/mL A = 0.17921 (flagged good), and the concentration that gives A = 1 in a 1 cm cell must be 156.2 µg/mL.

Those checks run with node tests/sci.mjs on every change to the site, together with every other science tool’s checks.

Check it yourself

On the Beer–Lambert calculator, in the A = ε l c card:

  1. Leave the first box on Absorbance and clear it, so the tool solves for absorbance.
  2. Put 2159.648 in ε, molar absorptivity (L/(mol·cm)) and 1 in Path length (cm).
  3. In Concentration type 5.9271 and set the unit to µmol/L (that is 2 µg/mL of famotidine). Type 337.435 in Molar mass (g/mol, optional).
  4. The big line reads Absorbance: 0.013, and under it Absorbance 0.0128 with the note “very low: noise and the blank matter a lot”, Transmittance 97.1 %, and Mass concentration 2.0 mg/L.
  5. Change Path length (cm) to 5: the absorbance becomes 0.0640, still low. Now set Concentration to 82.979 µmol/L (28 µg/mL) and the path back to 1: you get 0.1792, “in the most accurate range”.
  6. To find the concentration that fills the scale, clear Concentration, put 1 in the absorbance box and leave ε and the path length: the tool answers 463.0 µmol/L, and the mass concentration line shows 156.2 mg/L.

What researchers can use it for

  • Audit a published spectrophotometric method in a minute: multiply the Sandell sensitivity by ε and see whether you get the analyte’s molar mass.
  • Choose the cell before buying the instrument: try 1, 5 and 10 cm path lengths against your expected concentrations and pick the one that lands in A = 0.1 to 1.
  • Set a sensible calibration range: read off the concentrations for A = 0.1 and A = 1 at your ε, then dilute your standards to sit inside them.
  • Judge a detection limit: turn it into an absorbance and ask whether your instrument can really see that signal above its blank; fit the standards properly with the calibration curve, which gives the limit from the residual scatter.
  • Convert units safely: enter the molar mass once and the tool reports mol/L, mmol/L, µmol/L and mg/L together, which is where most assay arithmetic goes wrong.

The fine print

  • We used only the values printed in the paper’s text. Its calibration points, slope and intercept are in a figure and a table we did not reuse.
  • The 1 cm path length is the standard cuvette, but the paper does not state its cell, so every absorbance here is “per the path length shown”.
  • The ±0.002 A reading error is our assumption for an ordinary instrument, not a measurement of theirs.
  • Charge-transfer complexes drift as they age and with temperature. A Beer-Lambert check tells you nothing about that; it only tests the arithmetic of the numbers reported.
  • The paper’s own text gives recovery as 99.93 % in one place and 99.56 % in another. We report both rather than pick one.

Sources

  • Al Samarrai, S. Y., AlZubaidi, R. and Al-Ansari, N. (2024), Charge transfer complex-based spectrophotometric analysis of famotidine in pure and pharmaceutical dosage forms, Scientific Reports 14, doi:10.1038/s41598-024-54402-4. Licence CC BY 4.0 (Creative Commons Attribution), checked with Crossref metadata for the DOI, 2026-11-24. Values used here, all from the paper's own text: ortho-chloranil as the acceptor in an alkaline aqueous medium, absorbance read at 546 nm, Beer's law obeyed from 2 to 28 µg/mL, molar absorptivity 2,159.648 L/(mol·cm), Sandell sensitivity 0.15625 µg/cm², R² 0.9977, limit of detection 0.3024 µg/mL, limit of quantification 1.471 µg/mL, 2:1 reagent-to-drug stoichiometry, recovery 99.93 %.
  • Famotidine molecular formula C8H15N7O2S3; molar mass worked out from IUPAC/CIAAW standard atomic weights as used by fibub's Molar mass calculator.
  • Absorbance working range of about 0.1 to 1 on an ordinary spectrophotometer: Skoog, West, Holler and Crouch, Fundamentals of Analytical Chemistry (the usual textbook guidance on photometric error).