Calibrating an optical trap from a 2 ms fall: how much faster is it really?
A study calibrates optical tweezers from single trapping events and reports about 10 % precision per fall. We turn its corner frequency into a trap stiffness and ask how long an ordinary power-spectrum record needs to match it, with fibub's optical tweezers calibration tool.
An optical trap is a spring made of light, and before it can measure a force it has to be calibrated: you need its stiffness in piconewtons per nanometre. The standard way is to watch a trapped bead jiggle for a few seconds and fit the power spectrum of its position, which is a Lorentzian whose corner frequency gives the stiffness. A study in Micromachines asks whether you can do it from a single trapping event instead, by tracking a bead for about two milliseconds as it falls into the trap, and reports roughly 10 % precision per fall. This article converts the paper’s corner frequencies into stiffnesses and asks how long an ordinary equilibrium record has to be to match that precision, using the optical tweezers calibration tool. For the trapping focus itself see the Gaussian beam calculator, and for a closer look at any spectrum the FFT workbench.
The paper in brief
The authors trap 2 µm polystyrene beads with a 1064 nm laser and record what happens in the first couple of milliseconds of a trapping event, when the bead is still falling in. From that short trajectory they reconstruct the restoring force directly (the FORMA approach) and get the trap’s relaxation rate ω_c = 7834 ± 794 s⁻¹, against 6942 ± 870 s⁻¹ from the usual equilibrium power spectrum of the same trap, which they quote as about 118 pN µm⁻¹. Over five seconds their equilibrium estimate wanders between about 6900 and 7600 s⁻¹.
Their claims about precision are the interesting part: a single fall is within about 10 % of the true value 95 % of the time, the relative standard deviation across falls is below 6 % in experiment and 5 % in simulation, and a discretisation correction improves the recovered rate by about 25 %, giving 0.98 ± 0.08 of the true value from 2 ms sequences.
Checked with the tools
The paper’s raw traces are not published, so our records are simulated with the tool’s own exact discrete Ornstein-Uhlenbeck steps for a trapped bead, in a trap set to the paper’s corner frequency. Nothing here is the paper’s data.
Step 1: from the paper’s corner frequency to a stiffness. The tool’s Stokes drag for a 1 µm-radius bead in water at 25 °C (0.89 mPa·s, our assumption: the paper states neither the temperature nor the viscosity) is γ = 6πηr = 1.6776 × 10⁻⁸ N·s/m. The paper’s ω_c = 6942 s⁻¹ is a corner frequency of f_c = ω_c/2π = 1,104.9 Hz, so
k = γ ω_c = 116.5 pN µm⁻¹,
against the paper’s “about 118”. The 1.3 % gap is a viscosity choice, not a disagreement: 0.902 mPa·s, water at about 24.5 °C, gives exactly 118. Their quoted ±870 s⁻¹ is ±14.6 pN µm⁻¹, and their faster FORMA rate of 7834 s⁻¹ corresponds to 131.4 pN µm⁻¹, 12.8 % stiffer, which is within the two uncertainties but worth noticing.
| Check | Paper | Optical tweezers tool |
|---|---|---|
| Drag on the bead | not stated | 1.6776 × 10⁻⁸ N·s/m (2 µm bead, water at 25 °C) |
| Corner frequency, equilibrium | ω_c 6942 ± 870 s⁻¹ | f_c 1,104.9 Hz |
| Stiffness, equilibrium | about 118 pN µm⁻¹ | 116.5 pN µm⁻¹ (118.0 at 0.902 mPa·s) |
| Stiffness, falling particle | ω_c 7834 ± 794 s⁻¹ | 131.4 pN µm⁻¹ |
| Expected diffusion kT/γ | not stated | 0.245 µm²/s |
Step 2: what a one-second equilibrium record gives. Simulating one second at 100 kHz in that trap and running the tool’s pipeline (8,192-point segments, four bins averaged, aliased Lorentzian from 50 Hz to 20 kHz) returns f_c = 1,111.9 ± 21.4 Hz, a stiffness of 117.2 pN µm⁻¹, χ²/dof = 1.23, and an equipartition cross-check of 115.1 pN µm⁻¹ from a position scatter of 6.0 nm rms. So a second of data is worth about 2 %, well inside the paper’s ±12.5 % on the equilibrium value.
Step 3: how long does the spectrum need to match a 2 ms fall? We repeated the whole pipeline on 40 independent records at each of five lengths and measured the scatter of the fitted corner frequency:
| Record length | Segments averaged | Mean f_c | Scatter | Bias |
|---|---|---|---|---|
| 20 ms | 3 × 512 | 1,151 Hz | 15.0 % | +4.2 % |
| 50 ms | 2 × 2,048 | 1,139 Hz | 10.7 % | +3.1 % |
| 200 ms | 2 × 8,192 | 1,115 Hz | 6.2 % | +0.9 % |
| 1 s | 12 × 8,192 | 1,112 Hz | 1.8 % | +0.7 % |
| 5 s | 61 × 8,192 | 1,106 Hz | 0.7 % | +0.1 % |
So the paper’s 10 % per single fall is what a power spectrum achieves from about 50 ms of equilibrium data, and their sub-6 % repeat scatter needs about 200 ms. A 2 ms event is therefore worth roughly 25 to 100 times its own length in equilibrium recording: a real speed-up, and the reason the approach matters when a trap’s stiffness has to be measured once per bead rather than once per day.
The same table shows something the paper does not discuss: short spectra are biased high (+4.2 % at 20 ms), so hurrying the standard method makes a trap look stiffer than it is, which is the opposite direction to the roughly 10 % underestimate they warn about for equilibrium calibration.
Step 4: do not skip the aliasing correction. The paper is careful to fit an aliased Lorentzian, the exact spectrum of a sampled signal. We checked what that is worth. At 100 kHz the plain Lorentzian reads 1,166.7 Hz against the aliased 1,111.9 Hz, 4.9 % high. Sample the same trap at 5 kHz, a frame rate a fast camera might give, and the plain fit returns 1,663.5 Hz against a true 1,104.9: 51 % high, which is a 51 % error in the stiffness. The aliased model gets 1,096.7 Hz, 0.7 % low.
How we made the graphs
The charts and every number come from scripts/blog-figures/optical-tweezers-corner-frequency-fast-calibration.mjs in fibub’s public code. It calls drag (Stokes drag with Faxén’s correction), simulateBead (exact discrete Ornstein-Uhlenbeck steps), psd and block (the averaged periodogram), fitPsd (the analytical start plus Levenberg-Marquardt with Berg-Sørensen weights), lorentz, lorentzAliased and equipartition from src/lib/sci/phtweezers.ts, the maths behind the optical tweezers calibration tool.
The corner frequencies, the bead diameter and the precision figures are from the paper. The water viscosity (0.89 mPa·s), the temperature (25 °C), the 100 kHz sample rate, the record lengths and the 40 random seeds are our assumptions, and every record is simulated.
How we test the maths
On every change to fibub, tests/sci.mjs runs tests/sci-tools/optical-tweezers.mjs, which pins this tool against SciPy:
- the averaged periodogram against
scipy.signal.welch(boxcar window, no overlap) on the same record, including the segment count and bin frequencies; - the aliased and plain Lorentzian fits against
scipy.optimize.curve_fitwith the same reweighting: f_c = 379.993524 Hz and D = 491,613.14 nm²/s for the aliased fit on the tool’s example record; - equipartition against a NumPy variance, 205.9344 nm², giving 1.99889 × 10⁻⁵ N/m;
- that the fit recovers the simulated trap within two standard errors, with χ²/dof below 1.5.
- New with this article: a trap at the paper’s corner frequency must give a drag of 1.6776 × 10⁻⁸ N·s/m and a stiffness of 116.5 pN µm⁻¹ for a 2 µm bead in water at 25 °C, and a simulated one-second record at 100 kHz must return f_c = 1,111.9 Hz with the aliased model and 1,166.7 Hz without it.
Check it yourself
On the optical tweezers calibration page:
- Press Use the example. It loads a simulated record of a 1 µm bead sampled at 10 kHz, and reads Trap stiffness 0.02003 pN/nm at a Corner frequency fc of 380.0 ± 7.2 Hz, with Drag coefficient γ 8.388 × 10⁻⁹ N·s/m and an equipartition value of 0.01999 pN/nm.
- Change Model to Plain Lorentzian. The corner frequency jumps to 467.0 Hz and the stiffness to 0.02461 pN/nm, 22.9 % too stiff, because at 10 kHz the aliasing is not negligible. Change it back.
- Type Bead diameter (µm) 2: the drag doubles to 1.6776 × 10⁻⁸ N·s/m and the stiffness to 0.04005 pN/nm, while the corner frequency does not move. That drag is the one the paper’s bead has.
- Multiply it by the paper’s ω_c: k = 2πγf_c with f_c = 1,104.9 Hz gives 0.1165 pN/nm, the 116.5 pN µm⁻¹ above.
- Set Fit to (Hz) to 500, well below the corner, and watch the standard error on f_c grow: the corner has to be inside the fit range.
- Type a Height of the bead centre above the surface (µm) of 1.5 for the 2 µm bead: Faxén’s correction raises the drag, and the stiffness with it, by about a third.
What researchers can use it for
- Convert any published corner frequency into a stiffness (and back) with the drag the tool works out from the bead, the liquid and the height above the glass.
- Decide how long to record. The scatter table above is reproducible for your own trap: set the stiffness, the sample rate and the length, and read off the precision.
- Check a camera-rate calibration. At a few kHz the aliasing correction is worth tens of percent, far more than most other error sources.
- Cross-check with equipartition, which fails differently (it is spoiled by drift and noise rather than by bandwidth), so agreement between the two is meaningful.
- Teach the fluctuation-dissipation theorem with something tangible: kT/γ is the diffusion the fit must return if the positions are calibrated.
The fine print
- Our records are simulated; the paper’s traces are not published. The simulation is an ideal trapped bead: no drift, no detector noise, no hydrodynamic memory, no laser fluctuation, so our scatter figures are a best case.
- The paper’s method (force reconstruction from a 2 ms fall) is not the tool’s method (a Lorentzian fitted to an equilibrium spectrum). We compare precision at equal information, not the algorithms themselves, and a 2 ms record is far too short for the tool’s spectral route.
- The temperature and viscosity are ours, so our 116.5 pN µm⁻¹ carries that choice; the paper’s own number is about 118.
- The paper is from 2021, so the fast-calibration literature has moved on since; the arithmetic of corner frequency to stiffness has not.
Sources
- Stilgoe, A. B., Armstrong, D. J. and Rubinsztein-Dunlop, H. (2021), Enhanced Signal-to-Noise and Fast Calibration of Optical Tweezers Using Single Trapping Events, Micromachines 12(5): 570, doi:10.3390/mi12050570. Licence CC BY 4.0 (taken from the Crossref record; the page's own rights metadata says CC BY 3.0), checked 2026-11-24; the Crossref record carries no retraction or correction. Values used here: 2 micrometre polystyrene bead (refractive index 1.57) in water, 1064 nm trapping laser; equilibrium aliased power-spectrum fit omega_c = 6942 +- 870 per second, about 118 pN per micrometre; falling-particle (FORMA) result omega_c = 7834 +- 794 per second; equilibrium range over 5 s about 6900 to 7600 per second; simulated recovery 0.98 +- 0.08 from 2 ms sequences; precision about 10 % for a single fall, relative standard deviation below 6 % in experiment and 5 % in simulation; tracking noise below 1 nm with about 8 nm low-frequency noise.
- Power spectrum calibration, the aliased Lorentzian for sampled data and the expected scatter of averaged spectra: Berg-Sorensen, K. and Flyvbjerg, H. (2004), Review of Scientific Instruments 75: 594. Stokes drag and equipartition: standard statistical physics. Water viscosity 0.89 mPa s at 25 C is a rounded value.
If you are an author or publisher and want a credit or a passage changed or removed, tell us and we will act quickly.