Noise, filtering and lock-in detection

Experimental methods · 7 questions and templates for IIT JAM Physics, in order from warm-up to top rank. Check each answer, then read the full solution.

Loading questions… (they need JavaScript; nothing you type is sent anywhere)

Written questions with solutions

The same questions as a list, so you can print them or read without the practice box. Templates (fresh numbers) appear only in the practice box above.

  1. MCQWarm-up

    An amplifier has a voltage gain of 20 dB. Its output voltage is larger than the input by a factor of
    1. 20
    2. 100
    3. 10
    4. 20
    Answer and solution

    Answer: C

    For voltages, dB =20log10⁡(V2/V1), so 20 dB is a factor of 10 in voltage (and 100 in power).

    Why people get this wrong: 100 is the power ratio; 10 log is for power, 20 log for voltage.

  2. MCQWarm-up

    White noise is measured through a filter whose bandwidth is then made 100 times wider. The rms noise voltage becomes larger by a factor of
    1. 100
    2. 10
    3. 1 (it does not change)
    4. 50
    Answer and solution

    Answer: B

    White noise has constant power per hertz, so the noise power is proportional to bandwidth and the rms voltage to its square root: 100=10.

    Why people get this wrong: The power grows 100 times; the rms voltage only 10 times.

  3. MCQStandard

    An RC low-pass filter has R=10 kΩ and C=10 nF. Its −3 dB frequency is
    1. 10 kHz
    2. 159 Hz
    3. 6.28 kHz
    4. 1.59 kHz
    Answer and solution

    Answer: D

    RC=104×10−8=10−4 s. fc=1/(2πRC)=1/(6.283×10−4)=1.59 kHz.

    Why people get this wrong: 1/RC=104 s−1 is an angular frequency; divide by 2π for hertz.

  4. NATStandard

    A resistor is cooled from 300 K to 4.0 K (its resistance unchanged). Its Johnson noise voltage in a fixed bandwidth falls by a factor of ______ (to two decimal places).
    Answer and solution

    Answer: 8.66 (±0.01)

    vn=4kBTRB∝T, so the factor is 300/4.0=75=8.66. That is why sensitive preamplifier inputs and bolometers are cooled.

    Why people get this wrong: The noise power falls 75 times, the rms voltage only 75 times.

  5. MCQStandard

    A sensitive measurement shows strong 50 Hz hum and its harmonics only when the instrument and its preamplifier are both plugged into the mains and joined by a coaxial cable. The likely cause and cure are
    1. a ground loop; ground the system at one point only, or use a differential or isolated input
    2. Johnson noise in the cable; cool the cable
    3. shot noise; increase the signal current
    4. aliasing; add an anti-aliasing filter
    Answer and solution

    Answer: A

    The two mains earths and the cable shield form a closed loop. Mains magnetic fields induce a current round it, and earth currents give a potential difference between the two grounds; either way a 50 Hz voltage appears along the shield, in series with the signal. Breaking the loop (single-point grounding, isolation) or rejecting the common-mode voltage (differential input) removes it. Random noise would not be locked to the mains frequency.

    Why people get this wrong: Interference at exactly the mains frequency and its harmonics is pickup, not intrinsic noise.