Transducers and detectors
Experimental methods · 77 questions and templates for GATE Physics, in order from warm-up to top rank. Check each answer, then read the full solution.
Loading questions… (they need JavaScript; nothing you type is sent anywhere)
Written questions with solutions
The same questions as a list, so you can print them or read without the practice box. Templates (fresh numbers) appear only in the practice box above.
-
MCQWarm-up
A thermocouple produces a voltage because- its resistance changes with temperature
- a current is passed through it and it heats up
- two junctions of dissimilar metals are at different temperatures
- the metal emits electrons when heated
Answer and solution
Answer: C
The Seebeck effect: a temperature gradient along a conductor drives charge carriers, and two different metals do so by different amounts. Closing the loop with two junctions at different temperatures leaves a net EMF, roughly proportional to the temperature difference.Why people get this wrong: A thermocouple needs no excitation current; resistance thermometers do.
-
MCQWarm-up
The standard platinum resistance thermometer called Pt100 has a resistance of- 0 Ω at 0 °C
- 100 Ω at 0 °C
- 100 Ω at 100 °C
- 100 Ω at 25 °C
Answer and solution
Answer: B
The name gives the resistance at the ice point: 100 Ω at 0 °C, rising by about 0.385 Ω per kelvin (α = 0.00385 K⁻¹), so about 138.5 Ω at 100 °C.Why people get this wrong: Thermistors are usually specified at 25 °C; Pt100 at 0 °C.
-
MCQWarm-up
The gauge factor of an ordinary metal-foil strain gauge is- about 2
- about 0.3
- about 100
- about 1000
Answer and solution
Answer: A
For a metal, stretching lengthens and thins the wire, giving , and the resistivity changes a little, bringing the total to about 2. Semiconductor gauges reach 100 or more through the piezoresistive effect.Why people get this wrong: 0.3 is the Poisson ratio, not the gauge factor.
-
MCQWarm-up
Which detector is the natural choice for light at 1550 nm (the fibre-optic window)?- A silicon photodiode
- A bialkali photomultiplier
- A cadmium sulphide photoresistor
- An InGaAs photodiode
Answer and solution
Answer: D
A 1550 nm photon carries eV. Silicon ( eV) stops near 1.1 μm, and bialkali photocathodes and CdS respond to visible light. InGaAs ( eV) responds out to about 1.7 μm.Why people get this wrong: A detector responds only to photons above its band gap (or work function).
-
NATWarm-up
A photodiode has quantum efficiency 0.80 at 800 nm. Its responsivity is ______ A/W (to three decimal places). Use eV nm.Answer and solution
Answer: 0.516 A/W (±0.003)
Responsivity . A photon at 800 nm carries eV, so one watt is photons per second, each giving 0.8 electrons: A/W.Why people get this wrong: At fixed quantum efficiency the responsivity grows with wavelength, because each watt carries more photons.
-
MCQWarm-up
A Pirani gauge measures pressure through- the ionisation of gas by electrons
- the compression of a sample of gas
- the deflection of a thin diaphragm
- the heat carried away from a hot wire by the gas
Answer and solution
Answer: D
A heated wire loses heat by conduction through the gas. In the range about to 10 mbar the gas conductivity depends on pressure, so the wire temperature (its resistance) gives the pressure. The reading depends on the gas species.Why people get this wrong: Ionisation gauges work at much lower pressures.
-
MCQWarm-up
Which of these transducers is self-generating (active), producing its electrical signal without any external excitation?- a metal-foil strain gauge
- a thermistor
- a piezoelectric crystal
- a linear variable differential transformer (LVDT)
Answer and solution
Answer: C
A piezoelectric crystal produces charge directly when stressed. A strain gauge and a thermistor only change resistance, which must be read with a current or bridge supply, and an LVDT needs an AC excitation on its primary. Thermocouples and photovoltaic cells are other self-generating sensors.Why people get this wrong: An LVDT has an output coil, but it delivers nothing until its primary is driven.
-
NATStandard
A type K thermocouple (take a constant 41 μV/°C) has its hot junction at 400 °C and its reference junction at 25 °C. The EMF is ______ mV (to one decimal place).Answer and solution
Answer: 15.4 mV (±0.05)
The EMF depends on the difference of the junction temperatures: mV mV.Why people get this wrong: Using 400 °C alone assumes the reference is at 0 °C.
-
MCQStandard
A Pt100 sensor ( K, linear) is at 80 °C. Its resistance is- 108.0 Ω
- 130.8 Ω
- 138.5 Ω
- 100.3 Ω
Answer and solution
Answer: B
Ω.Why people get this wrong: 100.3 Ω forgets to multiply by properly; 138.5 Ω is the value at 100 °C.
-
MCQStandard
A 120 Ω strain gauge with gauge factor 2.1 is stretched by a strain of 500 microstrain. Its change in resistance is- 0.126 Ω
- 1.26 Ω
- 0.0126 Ω
- 0.252 Ω
Answer and solution
Answer: A
Ω, about 0.1% of the gauge resistance: why a bridge is needed to see it.Why people get this wrong: One microstrain is ; 500 microstrain is .
-
MCQStandard
One arm of a Wheatstone bridge (all four arms equal, supply 10 V) is a strain gauge of gauge factor 2 under a strain of . The bridge output is about- 10 mV
- 20 mV
- 5 mV
- 2.5 mV
Answer and solution
Answer: C
For a small fractional change in one arm, . Here , so mV.Why people get this wrong: A quarter bridge gives a quarter of ; two active arms give half, four give all of it.
-
MCQStandard
A photomultiplier has 10 dynodes, each with secondary emission ratio 4. Its gain is about- 40
Answer and solution
Answer: D
Each dynode multiplies the electron number by , so the gain is .Why people get this wrong: The gains multiply, not add: is wrong.
-
MSQStandard
Which statements about high-purity germanium (HPGe) gamma-ray detectors are correct?- They are operated cooled, near liquid-nitrogen temperature
- Their energy resolution is far better than that of NaI(Tl) scintillators
- An electron-hole pair costs about 3 eV of deposited energy
- They are scintillators read out by photomultipliers
Answer and solution
Answer: A, B, C
Germanium has a small gap (0.67 eV), so it must be cooled to keep the leakage current low. About 2.96 eV per pair means hundreds of thousands of carriers per MeV and small statistical spread: about 0.2% FWHM at 1.33 MeV against about 6 to 7% for NaI(Tl). The charge is collected directly; no light is involved.Why people get this wrong: Semiconductor detectors collect charge directly; scintillators convert to light first.
-
MCQStandard
The pulse height from a Geiger-Müller counter- is proportional to the energy deposited
- is the same for every ionising event, whatever its energy
- is proportional to the square root of the energy
- is zero for electrons
Answer and solution
Answer: B
In the Geiger region one primary ion pair triggers an avalanche that spreads along the whole anode wire until the space charge stops it. Every discharge is the same size, so a GM counter counts events but cannot measure energy.Why people get this wrong: Ionisation chambers and proportional counters do give energy information.
-
MCQStandard
Compared with a platinum resistance thermometer, an NTC thermistor- is more linear and works to higher temperatures
- has a resistance that rises with temperature
- has a positive coefficient about ten times larger
- is far more sensitive near room temperature but strongly non-linear
Answer and solution
Answer: D
An NTC thermistor is a semiconductor: , so falls steeply as rises (about per kelvin near room temperature, ten times Pt). The exponential makes it non-linear and its range is limited, typically to below about 150 °C.Why people get this wrong: NTC means negative temperature coefficient.
-
MCQStandard
An oil diffusion pump is always backed by a rotary pump because- its vapour jet works only when the outlet pressure is already low (about 0.1 mbar or less)
- it cannot pump light gases
- it needs the rotary pump to cool its oil
- it pumps only when rotating
Answer and solution
Answer: A
The jet of oil vapour drives gas molecules towards the outlet by momentum transfer. Against atmospheric pressure the jet collapses; it needs a forevacuum below its critical backing pressure, supplied by a rotary (or scroll) pump.Why people get this wrong: Diffusion pumps have no moving parts at all.
-
NATStandard
A thermometer with time constant 4.0 s, initially at 20 °C, is plunged into a bath at 80 °C. Its reading after 6.0 s is ______ °C (to one decimal place).Answer and solution
Answer: 66.6 °C (±0.1)
A first-order sensor obeys , so °C.Why people get this wrong: After one time constant the sensor has covered 63%, not 50% or 100%, of the step.
-
MCQStandard
A NaI(Tl) detector has an energy resolution (FWHM/E) of at keV. If the resolution is set by photoelectron statistics alone, at keV it is aboutAnswer and solution
Answer: B
The number of photoelectrons is proportional to , so the relative spread goes as : . The absolute width (in keV) grows as .Why people get this wrong: Scaling as (giving ) treats the width in keV as constant.
-
NATStandard
An incremental optical encoder has lines per revolution and its two quadrature channels are decoded on every edge of both (4x decoding). Its angular resolution is ______ degrees.Answer and solution
Answer: 0.09 (±0.001)
Two channels apart (electrically) give four edges per line: counts per revolution, so . The order in which the two channels change also tells the direction of rotation.Why people get this wrong: Counting lines only () wastes the resolution of quadrature decoding.
-
MCQStandard
A small amount of a halogen or organic vapour is added to the gas of a Geiger-Müller tube. Its purpose is- to raise the gas gain
- to absorb gamma rays
- to make the tube sensitive to slow neutrons
- to stop each discharge so the tube can reset after a pulse
Answer and solution
Answer: D
After an avalanche, positive ions drifting to the cathode, and UV photons from the discharge, could free new electrons and restart the discharge endlessly. The quench molecules absorb the UV photons and take over the ions' charge, then dissociate harmlessly, so each event gives one pulse. Halogen quenchers recombine, so halogen tubes last much longer than organic-quenched ones.Why people get this wrong: The gain in a GM tube is already saturated; the quench gas controls the end of the discharge, not its size.
-
NATExam level
Tables for a type K thermocouple (reference 0 °C) give 12.209 mV at 300 °C and 1.000 mV at 25 °C. If the reference junction sits at 25 °C and the hot junction at 300 °C, the measured EMF is ______ mV (to three decimal places).Answer and solution
Answer: 11.209 mV (±0.002)
Law of intermediate temperatures: , so mV. Cold-junction compensation adds back the 1.000 mV before the table is used.Why people get this wrong: Subtract EMFs, not temperatures, when the EMF-temperature curve is not linear.
-
MCQExam level
A Pt100 is read with four wires (two carry the current, two sense the voltage) instead of two. The main gain is that- the self-heating disappears
- the sensor becomes linear
- the resistance of the connecting leads no longer adds to the reading
- the sensitivity doubles
Answer and solution
Answer: C
With two wires the meter sees ; 1 Ω of leads is a 2.6 °C error. With four, the sensing pair carries almost no current, so there is no drop along it and the voltage is measured right at the sensor.Why people get this wrong: Self-heating depends on the current, which is the same in both arrangements.
-
NATExam level
An NTC thermistor follows with kΩ at 298.15 K and K. At 50 °C its resistance is ______ kΩ (to two decimal places).Answer and solution
Answer: 3.59 kΩ (±0.01)
K. Times : . kΩ.Why people get this wrong: Use kelvin in the exponent.
-
MCQExam level
The responsivity of a silicon photodiode rises with wavelength through the visible, peaks near 900 nm, then falls sharply beyond about 1100 nm. The sharp fall is due to- photons below the band gap (1.12 eV) no longer creating electron-hole pairs
- absorption in the glass window
- the photons becoming too energetic
- the junction capacitance
Answer and solution
Answer: A
The rise follows (more photons per watt). The cut-off is the band gap: nm. Beyond it silicon is transparent.Why people get this wrong: The fall at short wavelengths (UV) comes from absorption near the surface, a different effect.
-
MCQExam level
A GM counter with a dead time of 200 μs (non-paralysable) records 500 counts/s. The true count rate is about- 450 s
- 500 s
- 556 s
- 600 s
Answer and solution
Answer: C
The counter is dead for a fraction of the time. True rate s.Why people get this wrong: Dead time loses counts, so the true rate is higher than the observed one.
-
MCQExam level
As the voltage on a gas-filled detector is raised, the collected charge per event passes through several regions. In which region is the charge larger than the primary ionisation yet still proportional to it?- Recombination region
- Ionisation-chamber region
- Geiger-Müller region
- Proportional region
Answer and solution
Answer: D
In the proportional region electrons gain enough energy near the anode to ionise further, giving a gas gain of to that is the same for every event, so the pulse is still proportional to the deposited energy. In the ionisation region the gain is 1; in the GM region all pulses are equal.Why people get this wrong: The ionisation chamber is proportional too, but without gain.
-
MCQExam level
A Hall probe is a semiconductor slab 0.10 mm thick with m carrying 1.0 mA. In a field of 0.50 T perpendicular to the slab, the Hall voltage is about- 3.1 μV
- 31 μV
- 0.31 mV
- 3.1 mV
Answer and solution
Answer: B
V μV.Why people get this wrong: Thin, lightly doped material gives the largest Hall voltage: that is why probes use semiconductors, not metals.
-
MSQExam level
A photodiode can be used with reverse bias (photoconductive mode) or at zero bias (photovoltaic mode). Compared with zero bias, reverse bias gives- a faster response, because the wider depletion layer lowers the capacitance
- a larger dark current
- a photocurrent proportional to the light power over a wide range
- the lowest noise for very weak, slowly varying light
Answer and solution
Answer: A, B, C
Reverse bias widens the depletion region: lower capacitance (faster) and quick carrier collection, with a linear photocurrent. But the bias drives a dark (leakage) current whose shot noise matters for faint light; at zero bias there is no dark current, so photovoltaic mode is preferred for weak, slow signals.Why people get this wrong: Speed and linearity favour reverse bias; ultimate low-light noise favours zero bias.
-
MCQExam level
Which statement about a linear variable differential transformer (LVDT) is correct?- It works with a DC excitation
- Its output amplitude grows with core displacement, and the phase relative to the excitation flips by 180° as the core passes the centre
- Its output is largest with the core at the centre
- It measures velocity rather than displacement
Answer and solution
Answer: B
Two secondaries wound in opposition pick up equal voltages with the core centred, so the output is zero. Moving the core increases one and decreases the other: amplitude ∝ displacement, sign shown by the phase. A phase-sensitive detector (like a lock-in) turns this into a signed DC voltage.Why people get this wrong: An amplitude-only meter cannot tell left from right; the phase carries the direction.
-
NATExam level
A 50 L chamber is pumped with a constant effective speed of 5.0 L/s, ignoring outgassing. The time to go from 1000 mbar to 0.10 mbar is ______ s (to the nearest whole number).Answer and solution
Answer: 92 s (±1)
gives , so s.Why people get this wrong: Each factor of 10 in pressure costs the same time, .