Ten times fewer samples, 1.8 times the error: the fourth-root rule for sensor rates

A 2025 study estimates elbow angle from one low-power inertial sensor and finds that dropping from 100 Hz to 10 Hz raises the error by 1.8 times. We show why 1.8 is the most it could ever be.

A wearable that has to last a day on a 100 mAh battery spends most of its energy reading its sensor and doing arithmetic. Read the sensor ten times less often and you save roughly ten times the energy — but how much accuracy does that cost? A 2025 paper in the Journal of Low Power Electronics and Applications answers it for a single inertial device strapped to the forearm, estimating elbow angle for home rehabilitation: dropping from 100 Hz to 10 Hz raised the error by a factor of 1.8. That number is not an accident of their hardware. This article uses the kalman filter playground to show that 1.8 is almost exactly the ceiling for any filter of this kind, and where the ceiling comes from. For the 2 Hz low-pass in front of it see the FIR/IIR filter designer, and for combining the error terms, the uncertainty propagation calculator.

The paper in brief

The authors put one SensorTile module — an LSM303AGR accelerometer, an LSM6DSL gyroscope and an STM32L476 Cortex-M4 — on the forearm and compared five ways of getting the elbow angle out of it: the accelerometer alone, raw gyroscope integration, a bias-corrected gyroscope, a complementary filter and a Kalman filter. The signals are low-pass filtered at 2 Hz, since rehabilitation movements are slow.

Their results are about what a cheap single-device setup can do: RMSE between 1.83° and 5.52° at 100 Hz, depending on the test configuration and how fast the arm moves. The lightest algorithm costs 58 processor cycles, and the heavier fusion algorithms at least twice that. The engineering conclusion is the trade-off: cutting the data rate from 100 Hz to 10 Hz cuts sensor and computation power by about ten and multiplies the error by about 1.8.

Checked with the tools

Everything here is a simulation and a steady-state calculation, not a reanalysis of the paper’s recordings. The point is to test whether the paper’s 1.8 is the expected size, and what sets it.

Step 1: the fourth-root law. Take the simplest sensible model of a slowly moving joint: the true angle wanders like a random walk of variance q per second, and every sample measures it with noise of variance R. The tool’s steadyGainConstant solves the steady-state Kalman equations for a sample interval Δt. When the per-step wandering is small compared with the sensor noise — qΔt ≪ R, true for a slow joint and a 100 Hz sensor — the answer simplifies to

error ≈ (q Δt R)^(1/4)

so the error falls as the fourth root of the data rate. Ten times more samples buys 10^(1/4) = 1.778 times less error, no more. That is the paper’s 1.8.

Angle error against sensor data rateAngle error against sensor data rate0246850100150200Data rate (Hz)Steady-state error (degrees)one raw sample, 8°Kalman steady statefourth-root lawthe paper’s two ratesfibub.com
Figure 1. The steady-state error of a constant-level Kalman filter against the sensor data rate, for 8° of sample noise and a process noise chosen to put the 100 Hz error at the paper's 1.83°. The dashed line is the fourth-root law; the two dots are the paper's two data rates.

Step 2: putting the paper’s numbers in. Take the sensor noise on one accelerometer-derived angle sample as 8° (our assumption: the paper gives filtered errors, not raw sample noise). Solving for the process noise that puts the 100 Hz steady-state error at the paper’s best figure of 1.83° gives q = 18.5 deg² per second, that is an angle that wanders by about 4.3° in a second — a slow, deliberate rehabilitation movement. At that setting the tool’s exact steady state, with no small-qΔt approximation, gives:

Paper Tool (steady state)
Error at 100 Hz 1.83° (best case) 1.830° (fitted)
Error at 10 Hz 1.8 × the 100 Hz value 3.161°, that is 1.727 ×
Ceiling for a tenfold rate cut — 10^(1/4) = 1.778 ×

Step 3: the same thing from a simulation. Running the real filter rather than the formula, with the tool’s seeded simulator over one minute of data (6001 samples at 100 Hz, 601 at 10 Hz) and the filter’s Q and R matched to the truth, the measured RMSE averaged over 20 seeds is 1.85° at 100 Hz and 3.158° at 10 Hz, a ratio of 1.707. The raw measurements have an RMS error of about 8° at either rate, so the filter is doing the work: at 100 Hz it is averaging roughly 20 samples’ worth of noise.

Step 4: why 1.8 is a ceiling, not a coincidence. The fourth-root law only holds while the sensor noise dominates. Figure 2 sweeps the process noise q across six orders of magnitude and shows the penalty for a tenfold rate cut. It rises towards 1.778 for a slow-moving joint and falls towards 1.0 for a fast one, because a filter tracking a fast-wandering angle has to believe each new sample anyway and gains little from having more of them. There is no setting in which cutting the rate tenfold costs more than 1.778 times the error.

Cost of a tenfold slower sensorCost of a tenfold slower sensor11.21.41.61.80.010.1110100100010000Process noise q (deg² per s)Error at 10 Hz ÷ error at 100 Hzceiling 10^(1/4) = 1.778the paper measured 1.8q = 18.491fibub.com
Figure 2. How much a tenfold drop in data rate costs, as the angle's own wandering grows. The penalty can never exceed 10^(1/4) = 1.778, which is where the paper's measured factor of 1.8 sits.

That makes the paper’s measured 1.8 informative in a way its authors do not claim: it says their error at 100 Hz is almost entirely sensor noise being averaged away, not drift being tracked. If their error had been dominated by fast arm motion or gyroscope bias, the same rate cut would have cost much less than 1.8, and 10 Hz would have been nearly free. The flip side is that buying accuracy by sampling faster is expensive: going from 10 Hz to 100 Hz costs ten times the power for 1.8 times less error, and reaching 1° would need about 600 Hz.

How we made the graphs

The charts and every number come from scripts/blog-figures/kalman-filter-sampling-rate-elbow-angle-imu-fourth-root-law.mjs in fibub’s public code. It calls steadyGainConstant, simulate, kalman and summary from src/lib/sci/kalman.ts, the maths behind the kalman filter playground.

From the paper: the two data rates (100 Hz and 10 Hz), the 1.8 error factor, the tenfold power saving and the 1.83° best-case RMSE. Our assumptions: the constant-level (random-walk) model, 8° of noise on one angle sample, the process noise q = 18.5 deg²/s solved from the paper’s 1.83°, a one-minute record, and the filter’s Q and R set to the true values. The paper’s own recordings are not used; its rate-against-error curve is in a figure, not in its text.

How we test the maths

On every change to fibub, tests/sci.mjs runs tests/sci-tools/kalman-filter-playground.mjs, which pins the filter against an independent NumPy implementation written with full matrices:

  • On 40 made-up measurements at uneven times, every state, variance, innovation and innovation variance must match NumPy to 1 part in 10¹⁰, for the constant-level model and two constant-velocity settings.
  • The steady-state gain must match NumPy’s iteration of the Riccati equation to 1 part in 10¹², and a long run of the filter must settle onto it.
  • With Q and R matched to the simulated truth, the mean normalised innovation squared must come out near 1 and the estimate must beat the raw measurements.
  • New with this article: for 8° sample noise and q = 18.491 deg²/s, the steady-state error must be 1.830° at a 10 ms step and 3.161° at a 100 ms step, and the ratio of the two must stay below 10^(1/4) for every process noise from 0.01 to 10 000 deg²/s.

Check it yourself

On the kalman filter playground page:

  1. In Measurements, leave Data on Simulate a noisy sensor and Model on Constant level (random walk).
  2. Set Steps to 6001, Time step Δt (s) to 0.01, True process noise q to 18.491, Sensor noise σ to 8, Start value to 90 and Random seed to 7.
  3. Press Set Q and R to the true values so the filter is honest about the data it was given.
  4. In Result, read the RMS error of the estimate: about 1.95°, against about 8.07° for the raw measurements, with the normalised innovation squared near 1.0.
  5. Now set Steps to 601 and Time step Δt (s) to 0.1, keeping everything else, and press Set Q and R to the true values again. The estimate’s RMS error rises to about 3.10° — the same one minute of arm movement, a tenth of the samples, about 1.6 times the error on this seed.
  6. Press New noise a few times at each setting. Single runs wander by a few tenths of a degree; the average over many seeds settles near 1.85° and 3.16°.
  7. Raise True process noise q to 5000 at both rates and repeat. The gap between the two rates nearly closes: a fast-wandering angle does not reward a fast sensor.

What researchers can use it for

  • Choose a sampling rate before building the hardware: the fourth-root law turns an accuracy target into a rate, and shows when a faster sensor is not worth the battery.
  • Sanity-check a reported trade-off: if a paper claims a tenfold rate cut cost 3 times the error, something other than this filter is going on — aliasing, a gyroscope integration drift, or a mismatched Q.
  • Diagnose what dominates your error: measure the error at two rates; a ratio near 10^(1/4) means sensor noise, a ratio near 1 means real motion or bias.
  • Tune Q and R honestly using the normalised innovation squared, which should sit near 1; a value far from 1 means the filter’s noise settings do not match the data.
  • Teach the estimation trade-off with one slider, which is hard to do with equations alone.

The fine print

  • The paper’s 1.8 is a measurement across its test configurations; our 1.73 and 1.71 are a steady-state calculation and a simulation for a model we chose. They agree in size, which is the claim, not to the last digit.
  • A real elbow is not a random walk, and a real inertial device adds gyroscope bias, axis misalignment and soft-tissue movement. The fourth-root law describes the noise-averaging part of the error only.
  • The paper low-pass filters at 2 Hz. Sampling at 10 Hz is still above the Nyquist rate for that, so aliasing should be small, but it is another reason a measured factor can exceed the filter’s own ceiling.
  • The 8° sample noise is our assumption. A different value changes q but not the ratio, which is what this article is about.
  • Power scaling with data rate is rarely exactly tenfold: radio, housekeeping and leakage do not scale with sampling.

Sources

  • Fourniol, M., Vauche, R., Rao, G., Watelain, E. and Kussener, E. (2025), Elbow Joint Angle Estimation Using a Low-Cost and Low-Power Single Inertial Device for Daily Home-Based Self-Rehabilitation, Journal of Low Power Electronics and Applications 15(2): 33, doi:10.3390/jlpea15020033. Licence CC BY, checked on the MDPI article page, 2026-11-24. Values used here: one SensorTile device on the forearm with an LSM303AGR accelerometer and LSM6DSL gyroscope on an STM32L476 Cortex-M4; accelerometer and gyroscope signals low-pass filtered at 2 Hz; five estimators compared, including a complementary filter and a Kalman filter; root-mean-square error between 1.83 and 5.52 degrees at a 100 Hz sensor data rate; reducing the data rate from 100 Hz to 10 Hz increased the RMSE by a factor of 1.8 while cutting sensor and computation power by about a factor of 10; the lightest algorithm needs 58 processor cycles.
  • Steady-state Kalman filter for a random walk observed in white noise: standard estimation theory (the scalar Riccati equation), as in Kalman (1960) and estimation textbooks.

If you are an author or publisher and want a credit or a passage changed or removed, tell us and we will act quickly.